The PZ10 series nozzle flow meter is a low-flow measurement instrument based on the principle of magnetic induction. Fluid flows in from the inlet and acts on the sensor, causing its rotational speed to be proportional to the flow rate. The magnetic sensor detects the number of rotations within the chamber and emits a square wave pulse signal. The integrated model features a built-in display panel for direct reading of flow data. The pulse model outputs square wave pulses, making it suitable for connection to higher-level control systems such as PLCs, batch controllers, and counters.
Suitable for precise measurement of small flow rates, with good long-term operational consistency.
It can be combined with solenoid valves, PLCs/batch controllers to achieve quantitative supply of tap water or feed water.
PP/PVDF is optional, and the choice depends on the corrosion resistance of the liquid being measured.
The fluid flows in from the inlet and onto the sensor, causing it to rotate in proportion to the flow rate. The number of rotations within the chamber can be detected by the magnetic sensor, which then emits a square wave pulse signal. The higher the flow rate, the faster the sensor rotates; the pulse frequency is directly proportional to the flow rate, and the instantaneous flow rate and cumulative flow rate can be calculated by counting the pulses.
Proper installation is key to ensuring accurate measurement and long-term stable operation of the flow meter. Please follow the installation instructions below:
Note:Important points before installation: Before connecting to the pipeline, be sure to connect the electrical circuit first and test whether the signal is properly matched (you can manually turn the sensor or blow into the flow meter to make the sensor rotate and check the pulse signal).
Before installation, thoroughly clean the pipeline and install a filter (a 200-mesh or finer filter is recommended) upstream of the flow meter to prevent debris from entering the flow meter.
The flow meter should be installed horizontally, with the cover perpendicular to the horizontal plane. For the measurement of ultra-low flow rates, the recommended installation is bottom inlet and top outlet.
The arrow on the flow meter housing should point in the same direction as the flow of liquid in the pipe to be installed.
For pipes with continuous flow, the horizontal pipe where the flow meter is installed should be equipped with a bypass valve to facilitate regular cleaning and maintenance.
The following are common faults and solutions for the PZ10 series nozzle flow meter:
| Fault Symptoms | Possible Causes | Solution |
|---|---|---|
| The sensor does not rotate | Foreign objects are blocking the sensor, causing it to be jammed by debris. | Remove the meter, clean the sensor, and install a filter in the pipeline. |
| The sensor does not rotate | Clogged pipe filter | Clean the filter |
| Decreased flow rate (increased resistance) | Filter partially clogged | Clean the filter |
| Decreased flow rate (increased resistance) | Liquid viscosity is too high | Do not exceed the viscosity specified by the flow meter |
| The flow error is too large. | The actual flow rate range is lower or higher than the available range | Do not use beyond the limits |
| Excessive flow error | The liquid contains air | Avoid air in the pipeline |
| No pulse signal from the meter | Circuit malfunction | Replace the circuit |
| No pulse signal from the meter | Sensor jammed | The sensor and chamber need to be cleaned |