Learn about the latest news, industry updates and technical progress of Shanghai Cixi Instrument Co., Ltd.
The miniature Coriolis mass flow meter (DN1–DN15) is a direct mass flow meter developed specifically for low-flow, high-pressure applications. The sensor adopts a dual U-shaped vibrating tube structure. According to the Coriolis principle, a phase difference proportional to the mass flow rate is generated when the fluid passes through. The transmitter uses a high-precision DSP to calculate and directly obtain the mass flow rate, fluid density, and temperature. The measurement results are not affected by the viscosity, density, or pressure distribution of the fluid, making it particularly suitable for harsh media such as hydrogen, CNG, LNG, and chemical additives.
With a wide turndown ratio of up to 100:1 and the ability to measure flow rates as low as 0.2 kg/min, it meets the requirements for extremely low flow measurement and control in laboratories, fine chemical industries, and hydrogen refueling applications.
The CMASS003 has a maximum pressure resistance of 90 MPa and has passed a 1.5x pressure test, ensuring the safety and reliability of high-pressure hydrogen and natural gas dispensers.
Simultaneously outputs mass flow rate, volume flow rate, density, temperature, and concentration (calculated), reducing the number of on-site instruments and lowering integration costs.
There are no moving parts, no seal leakage points, the inner wall of the measuring tube is smooth, there is no additional pressure loss, and it is suitable for long-term continuous operation.
Coriolis mass flowmeters utilize the Coriolis effect generated when fluid flows through a vibrating tube. When the measuring tube vibrates at its natural frequency, the fluid flowing inside the tube generates a Coriolis force proportional to the mass flow rate, causing the vibrating tube to twist. The mass flow rate can be accurately calculated by detecting the phase difference between the inlet and outlet sides. At the same time, the resonant frequency of the vibrating tube is related to the density of the medium, allowing the density parameter to be obtained simultaneously. The CMASS series uses a DSP+ARM dual-core transmitter, which provides fast and stable signal processing and strong anti-interference capabilities.