What should be done if the accuracy of an oval gear flow meter decreases after long-term use?

2026-09-20 10:53:19
views
Reading time: about 5 minutes

Q:What should be done if the accuracy of an oval gear flow meter decreases after long-term use? What are the reasons for this? How can it be resolved and prevented?

A:A decrease in accuracy after prolonged use of an oval gear flow meter isa normal phenomenon.The main reasons areGear wear, increased bearing clearance, seal aging, corrosion or fouling of the metering chamber, which leads to an increase in internal leakage. The solution approach consists of three steps:First, troubleshoot issues related to the medium and installation. Next, inspect the meter for internal wear. Finally, based on the extent of wear, choose whether to recalibrate, replace worn parts, or replace the entire meter.If the decrease is only slight, it can be restored through recalibration. If the gears or bearings are significantly worn, the parts need to be replaced. If the measuring chamber is severely corroded or deformed, it is recommended to return the meter to the factory for overhaul or replace it with a new one. GN Instruments provides gear flow meter maintenance, calibration, and parts replacement services to help restore accuracy.

I. Six Main Causes of Decreased Accuracy

CauseMechanismTypical Manifestations
Gear wearAfter long-term operation, the gear tooth surface wears, the backlash increases, and internal leakage increases.Flow readings gradually decrease, with more pronounced deviations at low flow rates.
Increased bearing clearanceBearing wear causes axial or radial gear misalignment, reducing the sealing performance of the metering chamberReadings are unstable, repeatability is poor, and there may be abnormal noise.
Seal agingAging, hardening, or swelling of the sealing ring, leading to internal or external leakageExternal leakage is visible, while internal leakage results in lower readings.
Corrosion or fouling of the metering chamberCorrosive media erode the cavity, or crystallization and scaling of the media alter the cavity dimensions.Accuracy continues to decline, and damage or deposits on the cavity surface can be seen upon disassembly and inspection
Medium changesChanges in medium viscosity, temperature, and cleanliness that exceed the original calibration conditionsSudden change in accuracy, coinciding with the time of change in medium operating conditions
Installation issuesPipeline vibration, stress, air bubbles, or filter failure affect measurement stability.Readings fluctuate significantly, correlating with changes in installation conditions.

II. Troubleshooting and Resolution Steps

Step 1: Check for external factors
First, check whether the medium has changed, whether the filter is defective, whether there are bubbles or vibrations in the piping, and whether the installation is loose. These factors are easy to troubleshoot. After eliminating them, check the meter itself.
Step 2: Check the zero point and repeatability
Check for zero drift under no-flow conditions, and then test repeatability under stable flow conditions. If the repeatability is poor, it is usually due to mechanical wear or air bubbles.
Step 3: Disassemble and inspect for internal wear
Disassemble the flow meter to inspect the gear tooth surfaces, bearing clearances, sealing conditions, and chamber surfaces. Minor wear can be repaired, while severe wear requires replacement of parts.
Step 4: Recalibration
If the mechanical components are in good condition and only the readings are inaccurate, they can be corrected by recalibration. Calibration requires the use of actual media or standard flow devices.
Step 5: Replace worn parts
Gears, bearings, and sealing rings are considered wear parts and should be replaced when worn. After replacement, recalibration is required to ensure that accuracy is restored.
Step 6: Return to the factory for major overhaul or replacement
When the chamber is severely corroded or deformed, or the gears are severely worn, it is recommended to return the meter to the factory for major overhaul or to directly replace it with a new meter. The repair cost may be close to the price of a new meter.

III. Recommendations for Handling Different Degrees of Wear

Degree of WearSymptomsRecommended actionsCost reference
Slight (accuracy deviation < 1%)Readings are slightly off, but repeatability is acceptable.Re-calibration will restore normal operation.Low
Moderate (accuracy deviation 1%–5%)Readings are low, and repeatability is poor.Replace the sealing ring and bearing, and recalibrateModerate
Severe (accuracy deviation > 5%)Readings are significantly lower than expected, there is loud noise, and there may be a leak.Replace gears, bearings, and seals, and return to the factory for major overhaulHigh
Severe (chamber corrosion/deformation)Cannot be restored by replacing partsReplace with a new meterHighest

4. How to prevent a decrease in accuracy?

Preventive measures:

  • Install a filter at the front end:To prevent solid particles from entering the metering chamber, a filtration precision of 10 μm or higher is recommended.

  • Regular inspection and cleaning:Establish a maintenance schedule based on the cleanliness of the medium and clean the metering chamber regularly.

  • Avoid overspeed operation:Excessive rotational speed will exacerbate gear wear. It is recommended to operate within the recommended speed range.

  • Control the medium temperature:Avoid extreme temperature fluctuations to prevent thermal expansion from causing changes in clearance.

  • Correct model selection:Select the appropriate material based on the corrosivity of the medium to avoid corrosion caused by material incompatibility.

  • Regular calibration:It is recommended to calibrate once every year or two years to promptly detect any changes in accuracy.

  • Elimination of air bubbles:For media containing air, an air separator should be installed upstream of the flow meter.

V. GN Instruments Maintenance and Calibration Services

In response to the issue of declining accuracy in oval gear flowmeters,GN Instruments Instrument Co., Ltd.can provide the following services:

Service ItemsContentApplicable situations
Re-calibrationRecalibrate the flow meter using a standard flow device to correct the reading deviationMechanical components are in good condition, with only reading deviations
Replacement of partsReplace wear parts such as gears, bearings, and sealing ringsGear wear, increased bearing clearance, seal aging
Return to factory for major overhaulComplete disassembly and inspection, cleaning, replacement of worn parts, and recalibrationModerate to severe wear
Trade-inTrade in the old meter for a discount and replace it with a new oneThe chamber is severely corroded or deformed, and the repair cost is close to that of a new meter.

Based on your flow meter model, years of use, medium characteristics, and the degree of accuracy deviation, our engineers will determine whether recalibration, replacement of parts, or replacement with a new meter is necessary, and will provide the corresponding maintenance and calibration services.