Q:Is the Pirani vacuum gauge significantly affected by the type of gas? Is correction necessary when measuring different gases?

A:It has a significant impact.The Pirani gauge is a thermal conductivity vacuum gauge. Its measurement principle is as follows: Heat is transferred to the surrounding gas through the heating filament. The higher the gas pressure and the more molecules there are, the faster the heat transfer, the lower the temperature of the heating filament, and the lower the electrical resistance. Different gases haveThe thermal conductivity varies greatlyTherefore, at the same pressure, the readings of the Pirani vacuum gauge will vary significantly depending on the type of gas. Pirani gauges are typically calibrated usingnitrogen (N₂) or airAs a calibration reference, when measuring other gases such as helium, hydrogen, and argon, it is necessary to usegas correction factorfor conversion, or recalibrate for that gas. If the type of gas changes frequently or is unknown, it is recommended to use aCapacitance diaphragm vacuum gauge, which directly measures pressure regardless of the type of gas.

1. Why are Pirani vacuum gauges greatly affected by the type of gas?

The sensor of the Pirani vacuum gauge is a heated filament that conducts heat through gas molecules. Different gases have different molecular structures, molecular weights, and thermal conductivities, resulting in significant differences in heat transfer efficiency. The table below lists the relative thermal conductivities of common gases at room temperature (with nitrogen as 1.00):

GasRelative thermal conductivity (N₂=1.00)Pirani Reading TrendCorrection Factor Reference
Nitrogen (N₂)1.00 (reference)Reference1.00
Air≈1.00Near baseline≈1.00
Helium (He)≈5.7Reading too low (heat transfer too fast)≈0.18
Hydrogen (H₂)≈7.0Reading is too low≈0.14
Argon (Ar)≈0.67Reading is too high (heat transfer is too slow)≈1.50
Carbon dioxide (CO₂)≈0.62Reading is too high≈1.60
Sulfur hexafluoride (SF₆)≈0.35Readings are significantly higher than expected≈2.80

It can be seen that the thermal conductivity of helium and hydrogen is much higher than that of nitrogen, and the readings measured by the Pirani vacuum gauge will be significantly lower; whereas for gases with low thermal conductivity such as argon, carbon dioxide, and sulfur hexafluoride, the readings will be significantly higher. If no correction is made, the measurement error may reachseveral times or even an order of magnitude。

II. How to deal with the influence of gas types?

Use gas correction factors
Depending on the actual type of gas, multiply the Pirani vacuum gauge reading by the corresponding correction factor to obtain the true pressure. Most vacuum gauge controllers have built-in correction factors for common gases.
Re-calibration for gases
If a single gas is to be measured over a long period of time, the manufacturer can be asked to recalibrate the Pirani vacuum gauge for that gas so that the readings correspond directly to the target gas.
Use a compound vacuum gauge
Compound vacuum gauges (Pirani + ionization gauges) are still affected by gases in the low vacuum range, but the gas type can be switched via the controller to reduce manual conversion.
Switch to a capacitance diaphragm vacuum gauge
Capacitance diaphragm vacuum gauges directly measure pressure, regardless of the type of gas. This is the most reliable solution when the type of gas changes frequently or is unknown.
Exercise caution with mixed gases
The thermal conductivity of mixed gases depends on the proportion of each component, and the correction factor is difficult to determine accurately. It is recommended to switch to a vacuum gauge that is not affected by the type of gas.
Semiconductor Process Precautions
In semiconductor processes, gas types are frequently switched. The Pirani vacuum gauge needs to be used in conjunction with gas correction, or a capacitance diaphragm gauge should be selected to ensure accurate pressure control.

III. Examples of Correction Factor Usage

Correction factor calculation formula:

PActual = PReading × K

  • PActual:Actual pressure value

  • PReading:Pirani vacuum gauge reading (calibrated with N₂)

  • K:Correction factor for the gas

Example:A Pirani vacuum gauge calibrated with N₂ measures argon gas with a reading of 10 Pa. From the table, the correction factor for argon, K ≈ 1.50, so the actual pressure ≈ 10 × 1.50 = 15 Pa. If no correction is applied, the error can be as high as 50%.

IV. Selection Recommendations

Operating ConditionsRecommended SolutionReason
Gas type is fixed and known (e.g., N₂, air)Pirani vacuum gaugeLow cost, readings are directly usable
Fixed gas type, but not N₂ (e.g., Ar, He)Pirani vacuum gauge + correction factor or calibration for the specific gasEconomically viable, requires manual conversion or custom calibration
Frequent switching of gas typesCapacitance diaphragm vacuum gauge or compound vacuum gauge with gas switchingAvoid frequent conversions and human error
Mixed gases or unknown gas compositionCapacitance diaphragm vacuum gaugeDirectly measures absolute pressure, regardless of gas type
Corrosive gasesCorrosion-resistant capacitance diaphragm vacuum gaugeThe Pirani heating filament is susceptible to corrosion, and a corrosion-resistant diaphragm can be used for the capacitance diaphragm.

Before selecting a Pirani vacuum gauge, the following must be confirmed:

  • Gas type:Whether it is a single gas or a gas mixture, and whether there are frequent switches.

  • Measurement range:Select a Pirani vacuum gauge with an appropriate measuring range to cover the required vacuum range.

  • Accuracy requirements:Whether gas correction is required, and whether high-precision measurement is required.

  • Corrosiveness:Whether the gas is corrosive and whether a corrosion-resistant vacuum gauge is required.

  • Interfaces and Signals:Flange interface, output signal type, and whether a controller is required.

  • Environmental Conditions:The effects of temperature, humidity, vibration, etc., on measurements.

V. GN Instruments Vacuum Gauge Product Solution

Regarding the impact of gas types on vacuum measurements,GN Instruments Instrumentation Co., Ltd.can offer the following product solutions:

Product SeriesMeasurement RangeGas influenceApplicable Scenarios
CX-PRN Pirani Vacuum Gauge10⁻¹~10⁵ PaAffected by gas type, correction requiredRough and low vacuum monitoring, scenarios with fixed or correctable gas types
CX-PRN14 cold cathode Compound Vacuum Gauge10⁻⁵~10⁵ PaThe low vacuum section is affected by gas, and the gas can be switched via the controller.Wide-range monitoring, scenarios with minimal variation in gas types
CFG-500 Capacitance Diaphragm Gauge1~10⁵ PaIndependent of gas typeFrequent changes in gas types, mixed gases, precision vacuum control

Engineers will recommend suitable vacuum gauge models and configuration solutions based on your actual gas types, measurement range, accuracy requirements, and operating conditions. For scenarios where the gas type is fixed and can be corrected, a Pirani vacuum gauge is recommended; for scenarios where the gas type changes frequently or involves mixed gases, a capacitance diaphragm vacuum gauge is recommended. If necessary, we can provide prototype testing services to verify the measurement performance under actual operating conditions.