Can the MFC's measuring range be changed? Does it need to be recalibrated after the change?

2026-09-18 11:10:47
views
Reading time: about 5 minutes

Q:Can the range of the MFC (Mass Flow Controller) be changed? Does it need to be recalibrated after the change?

A:The range of the MFCcan be changed, but this is a technical process that requires professional handling, and re-calibration is necessary after the change.Users typically cannot perform this operation themselves, as changing the flow range often involves hardware replacement and precise calibration. The flow range is determined by an internal mechanical component called"flow splitter" (also known as a laminar flow element)To change the flow range, the key is to change the flow split ratio, which usually requiresReplace the splitter valve portor make similar hardware adjustments. This operation will directly change the measurement reference of the sensor. Therefore, after replacing the component, the entire MFC must undergorecalibrationto establish a new "sensor signal-actual flow rate" correlation.

1. Why does changing the range require "hardware modification"?

The MFC's range is not a simple software parameter. Its full-scale value is determined by an internal mechanical component called the"flow divider" (also known as a laminar flow element)mechanical component called the "flow divider" (also known as the laminar flow element). The flow splitter divides the total flow in a fixed proportion through its specific channel dimensions, with one path passing through the sensor for measurement and the other serving as the main flow path. To change the flow range, the key is to change this split ratio.

The flow divider determines the measuring range
The size of the flow divider's channel determines the proportional relationship between the flow measured by the sensor and the total flow. Generally, the larger the measuring range, the larger the flow divider channel.
Replacing the flow divider valve port
Changing the measuring range usually requires replacing the flow divider valve port or similar hardware to change the flow division ratio. This is a hardware-level change.
Calibration is required.
After replacing the hardware, the correlation between the sensor signal and the actual flow rate is completely altered, and re-calibration is necessary to restore accuracy.
Users cannot perform this operation themselves.
As precision components and calibration equipment are involved, the risk of users disassembling or adjusting them on their own is extremely high. Such tasks must be performed by the original manufacturer or an authorized service center.

II. Correct Procedure and Risks for Changing the Measuring Range

Since hardware replacement and precision calibration are involved, the correct operating procedure is toContact the original manufacturer or its authorized service centerDisassembly or adjustment by the user is not only extremely difficult to perform but may also lead to the following risks:

RiskDescriptionConsequences
Permanent loss of accuracyCalibration is a precision task, and incorrect operation can completely invalidate the measurement and control accuracy of the MFC.Inaccurate flow control, inaccurate proportioning, and scrapping of product batches
Damage to core componentsThe internal sensors and flow diverters have a precise structure, and non-professional handling can easily cause irreversible physical damage.The MFC is completely scrapped and needs to be repurchased.
Loss of warranty coverageOnce disassembled or adjusted by the user, the original manufacturer's warranty service is usually voided.Subsequent repairs must be paid for out of pocket, resulting in a significant increase in costs.

III. Corresponding Adjustments After Changing the Flow Range

After the range has been changed and calibrated, if your flow display unit displays flow in absolute units (such as SCCM, SLM), then the display unit also needs to be adjusted accordingly. There are two ways to handle this:

Adjust the display unit
Change the internal range setting of the display unit to the new full-scale value. If the display unit does not support self-adjustment, contact its manufacturer.
Use a correction factor
Instead of adjusting the display unit, multiply the displayed flow rate reading by a correction factor. Correction factor = New full-scale value / Original full-scale value.
Percentage Mode
If the display is in percentage (%) mode, no adjustment is usually required, as 100% always corresponds to the new full scale.

Key conclusions:

  • Can be changed:The MFC range can be changed by replacing hardware such as the shunt.

  • Must be recalibrated:After modification, the device must be recalibrated by the original manufacturer or an authorized service center; otherwise, accuracy cannot be guaranteed.

  • Do not attempt to perform this operation yourself:Users making changes themselves carry a very high risk, which may damage core components and void the warranty.

  • The display unit needs to be synchronized:After the change, the display unit needs to be adjusted simultaneously or a correction factor needs to be applied.

IV. Summary and Recommendations

Changing the MFC range is aprofessional service.. If you find that the initial range selection is inappropriate, the safest and most reliable way is toto contact the supplier or the original manufacturerto seek support. It is not advisable to attempt to make changes on your own, as the risks and potential losses far outweigh the cost of seeking professional assistance.

If you are facing an MFC selection issue, please provide yourActual operating flow rate rangeandTypes of gases used,GN Instruments Instrument Co., Ltd.can help you determine the appropriate flow range. We offer products such as gas mass flow meters/controllers, high-precision gas mass flow meters/controllers, and customized explosion-proof mass flow meters/controllers, and we can also provide selection recommendations and calibration services.